Theory of energy harvesting from heartbeat including the effects of pleural cavity and respiration
Abstract
Self-powered implantable devices with flexible energy harvesters are of significant interest due to their potential to solve the problem of limited battery life and surgical replacement. The flexible electronic devices made of piezoelectric materials have been employed to harvest energy from the motion of biological organs. Experimental measurements show that the output voltage of the device mounted on porcine left ventricle in chest closed environment decreases significantly compared to the case of chest open. A restricted-space deformation model is proposed to predict the impeding effect of pleural cavity, surrounding tissues, as well as respiration on the efficiency of energy harvesting from heartbeat using flexible piezoelectric devices. The analytical solution is verified by comparing theoretical predictions to experimental measurements. A simple scaling law is established to analyse the intrinsic correlations between the normalized output power and the combined system parameters, i.e. the normalized permitted space and normalized electrical load. The results may provide guidelines for optimization of in vivo energy harvesting from heartbeat or the motions of other biological organs using flexible piezoelectric energy harvesters.
1. Introduction
Implantable devices, aiming to solve problems such as heart failure, obesity, blood pressure control, wound healing, neural-stimulation and drug delivery, have made significant progress in clinical applications in recent years [1–5]. Pacemakers and implantable cardioverter defibrillators (ICDs) are most commonly used among these devices mentioned above. Both pacemakers and ICDs need a replacement every several years for various reasons, e.g. end of the battery life [6]. This needs additional surgeries which may bring a notable complication risk, especially in the case of involving lead additions [7]. On the other hand, these surgeries and device replacements are costly and apt to bring additional injuries to patients. However, the development of currently available battery technology is much slower than the electronic devices, which limits the pace of improvement in design and functionality of implantable electronic devices.
Self-powered implantable devices provide a potential way to address the shortcomings of battery replacement problem of implantable devices with the breakthrough technology of energy harvesting from ambient environments [8–13]. The main mechanical energy scavenging strategies are to harvest energy from biomechanical motions of human organs, such as heart-beating, blood flow and breathing etc. Lead zirconate titanate (PZT) is the mostly used material in mechanical energy harvesters (MEHs) for its high piezoelectric coefficients, and flexible/stretchable electronics with PZT material were studied intensively by Huang's group [14–18]. Based on these mechanical strategies, Dagdeviren et al. demonstrated a conformal device to collect electric power from motions of porcine heart, lung or diaphragm based on a flexible PZT MEH [19]. Ansari et al. [20] designed a fan-folded structure consisting of bimorph PZT beams for powering biomedical devices and sensors inside the body. Other piezoelectric materials, e.g. zinc oxide (ZnO), PMN-PT or PVDF, were also resorted to fabricate flexible MEHs [21–23]. For example, people may generate electric power using a flexible PVDF piezoelectric generator from the pulsation of ascending aorta of a swine or employing a ZnO breathing-driven implanted triboelectric nanogenerator powered by the rat's diaphragm [24–26].
Most recently, Lu et al. conducted a set of experiments to invest energy harvesting from the porcine heartbeat using flexible PZT MEH [27], in which the output voltage was measured under various conditions, i.e. chest open, chest closed and awake from anaesthesia. When the device is mounted on the heart surface in a matter of covering from the apex of the left ventricle (LV) to the right ventricle, the peak-to-valley amplitude of the output voltage reduces from 1.91 V with chest open to 1.09 V with chest closed, i.e. a reduction of 50%, while the average power is 0.02 µW with chest open to 0.005 µW with chest closed, i.e. a reduction of 75%. To demonstrate the underlying mechanism of the reduction in energy harvesting in chest closed environment relative to chest open, Zhang et al. [28] proposed a nonlinear compressive spring model to account for the impeding effects of surrounding tissues on the energy harvesting efficiency of the electromagnetic harvesting device. They treated the surrounding tissues, including the pericardium serosa, pleura and lung, approximately as a homogeneous medium that posed impeding effects on the deformation of the device. That is, once deformed, the device will be impeded by the spring. However, the anatomy shows that the cavum is enveloped by the visceral and parietal layer of the pericardium and that a gap of 5–10 mm exists between the parietal layer and the pleura [29]. As a result, the device may deform freely at its early stage and will be impeded only if the deformation amplitude exceeds this gap. In addition, the back-and-forth deformation of the device will also vary with respiration as the chest is closed.
In this paper, we propose a restricted deformation space model to demonstrate this mechanism of energy harvesting efficiency reduction in chest closed environment based on an electromechanical analysis. An analytical solution for the output voltage and power has been derived and will be validated by comparing with experimental measurements. A simple scaling law will be established which reveals the dependence of the normalized electric outputs on the normalized permitted space and the normalized electrical load. The output power can be tuned and optimized according to this scaling law by choosing properly the material, geometric and circuit parameters of the energy harvesting system. The results can also provide design guidelines for similar MEH devices that are applied in other biological environments.
2. Mechanical analyses of the mechanical energy harvester device
The PZT MEH is a layered structure (figure 1a), where PI (thickness ∼ 1.2 µm, E = 8.55 GPa) serves as the adhesive layer, Pt (thickness ∼ 0.3 µm, E = 196 GPa) serves as the bottom electrode, then PZT ribbons (length ∼ 2.2 mm, width ∼ 100 µm, thickness ∼ 500 nm, E = 128 GPa) serve as the functional layer which can generate electric power under deformation, Au (thickness ∼ 0.2 µm, E = 97 GPa) and Cr (thickness ∼ 10 nm, E = 292 GPa) serve as the top electrode and PI (thickness ∼ 1.2 µm, E = 8.55 GPa) serves as the top protective layer. Lithography and wet etching methods were employed to form the stack pattern (10 PZT nano ribbons connected in parallel in one module and 3 × 4 moduli connected in series) [27]. Then the functional part was transfer-printed to polyimide thin film (thickness ∼ 50 µm, E = 2.83 GPa), which serves as the soft bottom substrate layer. At last, the device is connected to a flexible anisotropic conductive film cable (figure 1b). To investigate the in vivo biomechanical energy harvesting efficiency, we integrated the device on the heart surface of experimental swine (figure 1c). The swine in the experiment was handled with the permission from the Ethics Committee. When the porcine heart was beating, the flexible PZT MEH was forced to deform, and the output voltage of the device was measured by an AD/DA card and transmitted to a computer for further analyses.
Figure 1. (a) Schematic illustration of layered structure of the flexible PZT MEH. (b) Photograph of the MEH device connected with cables. (c) Photograph of the MEH device that is mounted on the surface of the heart with chest open. (d) Schematic diagram of the device deformation with constraints of surrounding tissues.

As the outer LV is a curved surface, the device, when sewed at the LV surface by its two ends, is subjected to bending deformation with large deflections (figure 1d) during the heart motion in experiment [27]. The PZT nano ribbons are much more compliant than the substrate, and, therefore, the contribution of the PZT nano ribbons to the bending and stretching rigidity of the device is neglected. During the cardiac systole and diastole process, the device is forced to deform periodically. Neglecting the initial curvature of the device due to the LV surface, the deformation of the device is governed by
According to anatomy, there exists a small space (Ac = 5–10 mm [29]) between the left lung and the heart, this may accommodate free deformation of the device when the deformation amplitude A < Ac. However, the lung will pose an impedance effect on the device as A ≥ Ac, and thus, the output voltage is reduced. Therefore, the boundary conditions for the governing equation (2.1) will depend on the relationship between the deformation amplitude A and the permitted space Ac.
(a) Free deformation (A < Ac)
At the free deformation stage (A ≤ Ac), the MEH device is only constrained at its sewed ends with an end-to-end displacement of ΔL. The sewed ends are regarded as in clamped condition (figure 1d) to satisfy
The general solutions for the governing differential equation (2.1) are
Upon incorporating the boundary conditions in equation (2.2), the non-zero parameters in (2.3) and (2.4) are obtained as , and . During the systole of the heart, the relative end-to-end displacement of the device ΔL/L0 (approx. 10−1) is much larger than the membrane strain (approx. 10−5); therefore, the displacements are derived as
(b) Constrained deformation (A ≥ Ac)
When undergoing a large deformation (A > Ac) during the systole of the heart, the MEH device will contact with the surrounding tissues at the centre part. Here, the adhesion effect between the MEH device and the surrounding tissues is neglected, then the contact length will be zero [32,33], that is, the device will have only the centre point contacting with the surround tissues. For simplicity, the deformation of the surrounding tissues is further ignored or the surrounding tissues are assumed rigid when compared with the MEH device, then the displacement and the slope at the device centre will keep at and . Combining these constraint conditions and the symmetric nature of the device () with the boundary conditions in equation (2.2), the non-zero unknown constants ci are determined by
The average strain of the PZT ribbons may be approximated as
Noted that we have λ = π when Ac = A, then the solution in equation (2.9) may be reduced to equation (2.6) for the case of A < Ac.
3. Electric outputs of the mechanical energy harvester device
According to the electromechanical coupling behaviour of piezoelectric materials and following the previous analyses [28], the governing equation for total output voltage Vtotal of the MEH device with an electrical resistance R is
Equations (3.2) and (2.9) or (2.6) indicate that the output voltage of the MEH device may be calculated once the real-time end-to-end displacement of the device ΔL(t) is known.
For the current energy harvesting from heartbeat by sewing the MEH device at the LV surface, the end-to-end displacement ΔL(t), according to the assumption of bullet-shape or ellipsoid-shape [34], may be determined by , where twall, Ddiastolic and D(t) are, respectively, the wall thickness, diastolic and the real-time minor axis diameter of the LV. The real-time minor axis diameter D(t) is related to the volume of the LV by [35]. As the real-time LV volume VLV of swine is not available, we will use the corresponding in vivo data of human heart [36] to approximate the porcine LV due to their similar cardiac physiology [37–39]. The main difference between the real-time LV volume VLV of human and swine is the cardiac period, which makes it necessary to scale the cardiac period of human heart to the porcine heart in experiment.
The current PZT MEH device in the experiment contains 12 moduli connected in series with each module containing 10 PZT nano ribbons (hp = 0.5 µm, Ap = 0.22 mm2) connected in parallel. The effective piezoelectric and dielectric constants of the PZT nano ribbon are −6.2 C m−2 and 2 × 10−8 C/(Vm) [40], respectively. Using the loading condition generated by the above analogy, the predicted real-time output voltages with chest open and closed (AC = 6.5 mm, figure 2b) exhibit a reduction of about 40% in the peak-to-valley voltage, which is generally consistent with experiments (figure 2a). This demonstrates the capability of the current theoretical model to reveal the underlying mechanism of reduction in voltage output by the MEH device in chest closed environment.
Figure 2. Output voltage of the MEH when the chest is open and closed. (a) Experimental output voltage. (b) Predicted output voltage based on the analytical model. (Online version in colour.)

The energy harvesting efficiency is the foremost concern that determines the feasibility of the MEH device for practical applications to power biointegrated or implanted microdevices. Normally, the effective output power is used to evaluate the efficiency of the MEH device, where is the root of mean squares of the output voltage, or termed as the effective voltage. Theoretical predictions exhibit that the effective output power Peff increases monotonically with the permitted space Ac until it exceeds a certain value that will not pose impeding effects on the device (figure 3). Correspondingly, the effective output power will have a larger reduction for smaller permitted space. The predicted output power is 0.021 µW for chest open and 0.005 µW for chest closed with a permitted space AC = 6.5 mm (with a reduction of 76%), which are identical to the experimental measurements, 0.02 and 0.005 µW with chest open and closed (with a reduction of 75%) [27].
Figure 3. The electric power Peff and its relative decrement against the permitted space AC. (Online version in colour.)

In the chest closed environment, respiration may induce elastic deformation or volume change of the lung [41], and, hence, the timely variation of the pleural cavity or permitted space AC. The porcine cardiac rate is about 120–140 beats min−1 [41], while the respiratory rate is about 20 times min−1 [42]. Sinusoidal expiratory flow is usually used for studying lung mechanical properties [43], then we assume , where AC0 = 6.5 mm is the average permitted space [29], while AC1 and ωr are, respectively, the amplitude and frequency of the resonant space. According to the typical statistics of normal adult lung during respiration, these two parameters may be obtained approximately as AC1 = 3.0 mm and ωr = 2.1 rad s−1 [43]. Theoretical results of the real-time output voltage again agree very well with experiments (figure 4), indicating that the current theoretical model is robust for predicting energy harvesting from heartbeat in practical living environment. The result is also inspiring for potential applications in respiratory monitoring, for example, to retrieve real-time respiratory rate and respiratory intensity.
Figure 4. Comparisons of the predicted output voltage including the effects of respiration with experiments. (Online version in colour.)

4. Scaling laws for energy harvesting
Multiple parameters, including material parameters (e.g. piezoelectric constant , dielectric constant , Young's modulus of the substrate Es and the effective modulus of the whole device ), geometrical parameters (e.g. length of the device L0, thickness and effective area of the PZT ribbon hp and Ap, thickness of the substrate hs, moment of inertia of the substrate Is, effective moment of inertia of the whole device , the distance from the centre of the PZT layer to the centre of the PZT MEH zp, and the permitted space AC), circuit parameters (e.g. number of PZT ribbons np connected in parallel in one module, number of energy harvesting moduli ns connected in series and electrical load of the circuit R) and loading parameters (e.g. the real-time end-to-end displacement ΔL, the period of heartbeat T and the respiration rate ωr), are contained in the output voltage Vtotal given in equation (3.2). This makes the investigation of the effects of each individual parameter on the energy harvesting efficiency intricate and complex. To investigate the intrinsic effects of these parameters on the energy harvesting efficiency from heartbeat, we will establish simple scaling laws to reveal the underlying physics and links between the electric outputs and various parameters.
To this end, we introduce the intrinsic scales of voltage , time or electrical load , and then the normalized real-time output voltage including the effects of limited space may be expressed as
Figures 5 and 6 present the dependence of the normalized effective voltage Veff/V0, the normalized effective current Ieff/I0 and the normalized effective power Peff/P0 on the normalized electrical load R/R0 in the energy harvesting circuit for various normalized permitted space AC/L0. The normalized output voltage Veff/V0 increases monotonically with R/R0 and tends to saturate for large electrical load with the limit corresponding to the case of open circuit voltage (figure 5). In comparison, the normalized output current Ieff/I0 decreases monotonically with R/R0 and approaches zero for large electrical load, corresponding to the case of open circuit current. When the electrical load is zero, the energy harvesting circuit is short connected, resulting in a zero voltage and large current. For both the two extreme cases, the effective output power is zero which indicates that the MEH device does not harvest any electrical energy. For a practical energy harvesting circuit, the output power Peff/P0 attains the maximum as R/R0 = 0.130 (figure 6), nearly identical to 1/2π for the energy harvesting from harmonic motions [44]. The scaling law presented in the form of figure 6 is helpful to optimize the efficiency of energy harvester by selecting properly the electric load for the circuit or other system parameters. For example, for an MEH device with a fixed total number of PZT ribbons and moduli (nsnp is fixed), there exists an optimal ratio of np/ns to deliver a maximal output power. The optimal output power may also be realized by designing the geometry of the PZT ribbon (a proper Ap/hp) once the volume Aphp of the ribbon is given (figure 6).
Figure 5. Scaling law for the normalized effective voltage and normalized effective current for chest open (AC/L0 = 0.32) and closed (AC/L0 = 0.13) against the normalized circuit load R/R0. (Online version in colour.)
Figure 6. Scaling law for the normalized output power Peff/P0 for chest open (AC/L0 = 0.32) and closed (AC/L0 = 0.13) against the normalized circuit load R/R0. (Online version in colour.)

The scaling laws (figures 5 and 6) also reveal the adverse effects of the limited permitted space AC/L0 on the efficiency of the MEH device. The smaller the permitted space is, the lower output power is achieved (figure 6). According to the physiology of heart motion, the resulted maximal end-to-end displacement of the device is [36], which can be substituted into the deflection amplitude in equation (2.5) leading to the maximal amplitude as Amax/L0 = 0.32. This indicates that the device will behave like the case of free deformation as the permitted space AC/L0 > 0.32. Thus, the normalized output power Peff/P0 keeps increasing until AC/L0 exceeds 0.32 where the power remains constant and is equal to the case of chest open (figure 7). The effective power produced by the energy harvesting system in the experiment (R/R0 = 0.192) is a little bit lower than the optimal case R/R0 = 0.130. The scaling law presented in the form of figure 7 may also provide guidelines for designing the MEH device to achieve as higher output power as possible. Combining the investigations on equation (2.9) and the curve in figure 7, the output power is inversely proportional to at least the fourth order of the device length L0, that is, shorter devices (smaller L0) may produce higher output power. Taking the device in experiment (R/R0 = 0.192, L0 = 5.0 cm), for example, the effective output power Peff may increase up to 2.5 times if the device length decreases by 20% (L0 = 4.0 cm). Even if the device is so short that the relative permitted space satisfies AC/L0 > 0.32, the output power will keep inversely proportional to the square of the device length L0. The scaling law also implies that the output power is proportional to the square of the bending rigidity ratio , effective piezoelectric constant and the distance zp of the PZT ribbons away from the neutral axis of the device.
Figure 7. Scaling law for the normalized output power Peff/P0 against the normalized permit space Ac/L0 for various normalized circuit loads R/R0. (Online version in colour.)

5. Conclusion
We have developed a restricted-space deformation model to predict the efficiency of energy harvesting from the heartbeat using flexible PZT devices. The model was validated by comparing analytical results to experimental measurements, which demonstrated its capability of revealing the underlying mechanism of reduction in the efficiency of energy harvesting in chest closed environment compared to the case of chest open. A simple scaling law for the energy harvesting from heartbeat is established in terms of material, geometric, circuit and heartbeat parameters. The normalized power output by the energy harvesting system only depends on two combined parameters, i.e. the normalized pleural cavity and the normalized electrical load of the circuit. With prescribed PZT ribbons and energy harvesting moduli, the most efficient means of increasing the energy harvesting efficiency is to select proper electrical load and decrease the device length. The results may serve as guidelines for optimization of energy harvesting from heartbeat or the motion of other biological organs.
Ethics
All in vivo experiments performed were approved by the Ethics Committee (Ethics Committee of Beijing Anzhen Hospital, Capital University of Medical Sciences) and in accordance with relevant guidelines and regulations. Experimental swine raised especially for cardiovascular research are selected for in vivo testing.
Data accessibility
The datasets supporting the conclusions are accessible in the article.
Authors' contributions
Y.Z. carried out the analytic modelling, analysed the data and drafted the manuscript; C.L. designed the analytic model, analysed the data, revised and finalized the manuscript. B.L. and X.F. provided the experimental data and photos.
Competing interests
We declare we have no competing interests.
Funding
C.L. acknowledges the financial supports by the National Natural Science Foundation of China at grant nos 11322216 and 11621062. X.F. acknowledges the supports by the National Natural Science Foundation of China at grant no. 11320101001. B.L. acknowledges the support of the National Natural Science Foundation of China at grant no. 11502126.
Acknowledgements
The authors thank the reviewers for their helpful comments and suggestions for improving this article.
Footnotes
References
- 1
Lohmeier TE, Iliescu R, Liu B, Henegar JR, Maricbilkan C, Irwin ED . 2012Systemic and renal-specific sympathoinhibition in obesity hypertension. Hypertension 59, 331–338. (doi:10.1161/hypertensionaha.111.185074) Crossref, PubMed, ISI, Google Scholar - 2
Kobayashi M, Massiello A, Karimov JH, Van Wagoner DR, Fukamachi K . 2013Cardiac autonomic nerve stimulation in the treatment of heart failure. Ann. Thorac. Surg. 96, 339–345. (doi:10.1016/j.athoracsur.2012.12.060) Crossref, PubMed, ISI, Google Scholar - 3
Potkay JA . 2008Long term, implantable blood pressure monitoring systems. Biomed. Microdevices 10, 379–392. (doi:10.1007/s10544-007-9146-3) Crossref, PubMed, ISI, Google Scholar - 4
Lavan DA, Mcguire T, Langer R . 2003Small-scale systems for in vivo drug delivery. Nat. Biotechnol. 21, 1184–1191. (doi:10.1038/nbt876) Crossref, PubMed, ISI, Google Scholar - 5
Rak-Hwan K 2012Flexible electronics: materials and designs for wirelessly powered implantable light-emitting systems. Small 8, 2812–2818. (doi:10.1002/smll.201200943) Crossref, PubMed, ISI, Google Scholar - 6
Mond HG, Proclemer A . 2011The 11th world survey of cardiac pacing and implantable cardioverter-defibrillators: calendar year 2009—a World Society of Arrhythmia's project. PACE 34, 1013–1027. (doi:10.1111/j.1540-8159.2011.03150.x) Crossref, Google Scholar - 7
Poole JE 2010Complication rates associated with pacemaker or implantable cardioverter-defibrillator generator replacements and upgrade procedures: results from the REPLACE registry. Circulation 122, 1553–1561. (doi:10.1161/circulationaha.110.976076) Crossref, PubMed, ISI, Google Scholar - 8
Kim KH, Lee KY, Seo JS, Kumar B, Kim SW . 2011Paper-based piezoelectric nanogenerators with high thermal stability. Small 7, 2577–2580. (doi:10.1002/smll.201100819) Crossref, PubMed, ISI, Google Scholar - 9
Bowen CR, Kim HA, Weaver PM, Dunn S . 2013Piezoelectric and ferroelectric materials and structures for energy harvesting applications. Energy Environ. Sci. 7, 25–44. (doi:10.1039/c3ee42454e) Crossref, ISI, Google Scholar - 10
Yang Y, Zhang H, Chen J, Lee S, Hou TC, Wang ZL . 2013Simultaneously harvesting mechanical and chemical energies by a hybrid cell for self-powered biosensors and personal electronics. Energy Environ. Sci. 6, 1744–1749. (doi:10.1039/c3ee40764k) Crossref, ISI, Google Scholar - 11
Lee JH, Lee KY, Kumar B, Tien NT, Lee NE, Kim SW . 2012Highly sensitive stretchable transparent piezoelectric nanogenerators. Energy Environ. Sci. 6, 169–175. (doi:10.1039/c2ee23530g) Crossref, ISI, Google Scholar - 12
Antoine GO, De LE, Michelin S . 2016Optimal energy harvesting from vortex-induced vibrations of cables. Proc. R. Soc. A 472, 20160583. (doi:10.1098/rspa.2016.0583) Link, Google Scholar - 13
Renzi E . 2017Hydroelectromechanical modelling of a piezoelectric wave energy converter. Proc. R. Soc. A 472, 20160715. (doi:10.1098/rspa.2016.0715) Link, Google Scholar - 14
Rogers JA, Someya T, Huang Y . 2010Materials and mechanics for stretchable electronics. Science 327, 1603–1607. (doi:10.1126/science.1182383) Crossref, PubMed, ISI, Google Scholar - 15
Ma Y, Xue Y, Jang KI, Feng X, Rogers JA, Huang Y . 2016Wrinkling of a stiff thin film bonded to a pre-strained, compliant substrate with finite thickness. Proc. R. Soc. A 472, 20160339. (doi:10.1098/rspa.2016.0339) Link, Google Scholar - 16
Khang D-Y, Jiang H, Huang Y, Rogers JA . 2006A stretchable form of single-crystal silicon for high-performance electronics on rubber substrates. Science 311, 208–212. (doi:10.1126/science.1121401) Crossref, PubMed, ISI, Google Scholar - 17
Kim DH 2008Materials and noncoplanar mesh designs for integrated circuits with linear elastic responses to extreme mechanical deformations. Proc. Natl Acad. Sci. USA 105, 18 675–18 680. (doi:10.1073/pnas.0807476105) Crossref, ISI, Google Scholar - 18
Zhang Y, Fu H, Su Y, Xu S, Cheng H, Fan JA, Hwang KC, Rogers JA, Huang Y . 2013Mechanics of ultra-stretchable self-similar serpentine interconnects. Acta Mater. 61, 7816–7827. (doi:10.1016/j.actamat.2013.09.020) Crossref, ISI, Google Scholar - 19
Dagdeviren C 2014Conformal piezoelectric energy harvesting and storage from motions of the heart, lung, and diaphragm. Proc. Natl Acad. Sci. USA 111, 1927–1932. (doi:10.1073/pnas.1317233111) Crossref, PubMed, ISI, Google Scholar - 20
Ansari MH, Karami MA . 2017Experimental investigation of fan-folded piezoelectric energy harvesters for powering pacemakers. Smart Mater. Struct. 26, 065001. (doi:10.1088/1361-665X/aa6cfd) Crossref, PubMed, ISI, Google Scholar - 21
Wang ZL, Song J . 2006Piezoelectric nanogenerators based on zinc oxide nanowire arrays. Science 312, 242–246. (doi:10.1126/science.1124005) Crossref, PubMed, ISI, Google Scholar - 22
Hwang GT 2014Self-powered cardiac pacemaker enabled by flexible single crystalline PMN-PT piezoelectric energy harvester. Adv. Mater. 26, 4880–4887. (doi:10.1002/adma.201400562) Crossref, PubMed, ISI, Google Scholar - 23
Chang C, Tran VH, Wang J, Fuh YK, Lin L . 2010Direct-write piezoelectric polymeric nanogenerator with high energy conversion efficiency. Nano Lett. 10, 726–731. (doi:10.1021/nl9040719) Crossref, PubMed, ISI, Google Scholar - 24
Zhang H 2015A flexible and implantable piezoelectric generator harvesting energy from the pulsation of ascending aorta: in vitro and in vivo studies. Nano Energy 12, 296–304. (doi:10.1016/j.nanoen.2014.12.038) Crossref, ISI, Google Scholar - 25
Zheng Q 2014In vivo powering of pacemaker by breathing-driven implanted triboelectric nanogenerator. Adv. Mater. 26, 5851–5856. (doi:10.1002/adma.201402064) Crossref, PubMed, ISI, Google Scholar - 26
Wang ZL, Chen J, Lin L . 2015Progress in triboelectric nanogenerators as a new energy technology and self-powered sensors. Energy Environ. Sci. 8, 2250–2282. (doi:10.1039/c5ee01532d) Crossref, ISI, Google Scholar - 27
Lu B 2015Ultra-flexible piezoelectric devices integrated with heart to harvest the biomechanical energy. Sci. Rep. 5, 16065. (doi:10.1038/srep16065) Crossref, PubMed, ISI, Google Scholar - 28
Zhang Y, Chen Y, Lu B, Lü C, Feng X . 2016Electromechanical modeling of energy harvesting from the motion of left ventricle in closed chest environment. J. Apll. Mech-T. ASME 83, 061007. (doi:10.1115/1.4032994) Crossref, ISI, Google Scholar - 29
- 30
Li Y, Song J, BoFang Zhang J . 2011Surface effects on the postbuckling of nanowires. J. Phys. D. Appl. Phys. 44, 425304. (doi:10.1088/0022-3727/44/42/425304) Crossref, ISI, Google Scholar - 31
Vella D, Bico J, Boudaoud A, Roman B, Reis PM . 2009The macroscopic delamination of thin films from elastic substrates. Proc. Natl Acad. Sci. USA 106, 10 901–10 906. (doi:10.1073/pnas.0902160106) Crossref, ISI, Google Scholar - 32
Yang F . 2003Contact deformation of a micromechanical structure. J. Micromech. Microeng. 14, 263–268. (doi:10.1088/0960-1317/14/2/014) Crossref, ISI, Google Scholar - 33
Lü C, Wu D, Chen W . 2011Surface effects on the jump-in instability of nanomechanical structures. IEEE Trans. Nanotechnol. 10, 962–967. (doi:10.1109/tnano.2010.2091422) Crossref, ISI, Google Scholar - 34
Lang RM 2005Recommendations for chamber quantification. J. Am. Soc. Echocardiogr. 18, 1440–1463. (doi:10.1016/j.echo.2005.10.005) Crossref, PubMed, ISI, Google Scholar - 35
Teichholz LE, Kreulen T, Herman MV, Gorlin R . 1976Problems in echocardiographic volume determinations: echocardiographic-angiographic correlations in the presence of absence of asynergy. Am. J. Cardiol. 37, 7–11. (doi:10.1016/0002-9149(76)90491-4) Crossref, PubMed, ISI, Google Scholar - 36
Corsi C, Saracino G, Sarti A, Lamberti C . 2002Left ventricular volume estimation for real-time three-dimensional echocardiography. IEEE Trans. Med. Imaging. 21, 1202–1208. (doi:10.1109/tmi.2002.804418) Crossref, PubMed, ISI, Google Scholar - 37
Hill AJ, Laske TG, Sigg DC, Skadsberg ND, Vincent SA, Soule CL, Gallagher WJ, Iaizzo PA . 2005In vitro studies of human hearts. Ann. Thorac. Surg. 79, 168–177. (doi:10.1016/j.athoracsur.2004.06.080) Crossref, PubMed, ISI, Google Scholar - 38
Sigg DC, Iaizzo PA . 2006In vivo versus in vitro comparison of swine cardiac performance: induction of cardiodepression with halothane. Eur. J. Pharmacol. 543, 97–107. (doi:10.1016/j.ejphar.2006.06.011) Crossref, PubMed, ISI, Google Scholar - 39
Mohiuddin MM, Reichart B, Byrne GW, Mcgregor CG . 2015Current status of pig heart xenotransplantation. Int. J. Surg. 23, 234–239. (doi:10.1016/j.ijsu.2015.08.038) Crossref, PubMed, ISI, Google Scholar - 40
Su Y, Dagdeviren C, Li R . 2015Measured output voltages of piezoelectric devices depend on the resistance of voltmeter. Adv. Funct. Mater. 25, 5320–5325. (doi:10.1002/adfm.201502280) Crossref, ISI, Google Scholar - 41
Liedtke AJ, Nellis S, Neely JR . 1978Effects of excess free fatty acids on mechanical and metabolic function in normal and ischemic myocardium in swine. Circ. Res. 43, 652–661. (doi:10.1161/01.RES.43.4.652) Crossref, PubMed, ISI, Google Scholar - 42
Heitman H, Hughes EH . 1949The effects of air temperature and relative humidity on the physiological well being of swine. J. Anim. Sci. 8, 171–181. (doi:10.2527/jas1949.82171x) Crossref, ISI, Google Scholar - 43
Bitzén U, Niklason L, Göransson I, Jonson B . 2010Measurement and mathematical modelling of elastic and resistive lung mechanical properties studied at sinusoidal expiratory flow. Clin. Physiol. Funct. Imaging, 30, 439–446. (doi:10.1111/j.1475-097X.2010.00963.x) Crossref, PubMed, ISI, Google Scholar - 44
Lu F, Lim SP, Lee HP . 2003Modeling and analysis of micro piezoelectric power generators for micro-electromechanical-systems applications. Smart Mater. Struct. 13, 57–63. (doi:10.1088/0964-1726/13/1/007) Crossref, ISI, Google Scholar


